Parameter / Equipment
Temperature controlled enclosure
- Type / Subtype
- General
- Accreditation program
- TISI 0261 — Site
- Service Location
- TISI 0261 — Site
4 Range Rows
Range
-5 °C to 50 °C
CMC / Expanded Uncertainty
0.34 °C
Range
>50 °C to 100 °C
CMC / Expanded Uncertainty
0.40 °C
Range
>100 °C to 150 °C
CMC / Expanded Uncertainty
0.65 °C
Range
>150 °C to 200 °C
CMC / Expanded Uncertainty
0.80 °C
Evidence DetailsShowHide
- Source Row 1
- -5 °C to 50 °C / 0.34 °C
- Method
- TLAS G-20
- Source Group
- Official source group
- Source Document
- TISI 0261 certificate 22-LB0022, Calibration 0261, Issue No. 03
- Source Page
- 16
- Medium / Condition
- Service Location
- TISI 0261 — Site
- Reference Equipment
- Source Row 2
- >50 °C to 100 °C / 0.40 °C
- Method
- TLAS G-20
- Source Group
- Official source group
- Source Document
- TISI 0261 certificate 22-LB0022, Calibration 0261, Issue No. 03
- Source Page
- 16
- Medium / Condition
- Service Location
- TISI 0261 — Site
- Reference Equipment
- Source Row 3
- >100 °C to 150 °C / 0.65 °C
- Method
- TLAS G-20
- Source Group
- Official source group
- Source Document
- TISI 0261 certificate 22-LB0022, Calibration 0261, Issue No. 03
- Source Page
- 16
- Medium / Condition
- Service Location
- TISI 0261 — Site
- Reference Equipment
- Source Row 4
- >150 °C to 200 °C / 0.80 °C
- Method
- TLAS G-20
- Source Group
- Official source group
- Source Document
- TISI 0261 certificate 22-LB0022, Calibration 0261, Issue No. 03
- Source Page
- 16
- Medium / Condition
- Service Location
- TISI 0261 — Site
- Reference Equipment