pH measurement fundamentals
pH buffer adjustment versus laboratory calibration

The instrument's Cal menu may adjust slope and offset. A successful screen message and a laboratory calibration report serve different evidence needs.
Read what the Cal routine actually does
Manufacturer calibration routines may also adjust the relationship between electrode signal and displayed pH. Record buffers, temperature and setting changes under the model's instructions.
References: Hanna Instruments: How do I calibrate my pH electrode? · METTLER TOLEDO: pH Theory Guide
Define the calibration system and results
Agree whether work covers the readout, electrode or complete system. Calibration reports results and associated uncertainty under stated conditions; it does not automatically include adjustment or a pass statement. An mV readout check is not a complete electrode assessment in solution.
Retain evidence before and after adjustment
Request as-found results before changing settings. Agree adjustment, subsequent results and any acceptance criteria and decision rule separately. Routine care and intermediate checks remain necessary according to your procedure.
FAQ
Does Cal success replace a calibration report?
Not automatically. Check the required instrument identity, points, conditions, results and uncertainty against your organization's requirements.